Magnetic Separation Tube

Short Description:

The XCGS-type magnetic separation tube is the Davis analytical tube, which is suitable for laboratories in mining, metallurgy, geology, etc. It is used to determine the content of strong magnetic components and is a device for inspecting these ores during the separation process.

1.Principle:
XCGS - type magnetic tube is in C type electromagnet between two extremes, equipped with glass tube to do reciprocating movement and swing, by sorting the sample through a magnetic field, the magnetic part is attached to the wall near the non-magnetic parts in the mechanical movement is wash and discharge water.
2.Type structure introduction:
XCGS - magnetic separation tube is composed of host and excitation power supply. The host as shown in figure (a). Installed on the base of the bracket (4) (5) is by core C (1) and the field winding (3) fixed support, in the extreme with magnetic flux hoop (6) so as to hold bracket (7). The glass tube (2) is embedded in the circular ring (11) and placed between the magnetic poles through the sliding frame (10). The motor (8) makes it swing through the reduction mechanism (9) with the help of the crank (12). Glass tube device with horizontal line to 0 °, 40 ° Angle. The reciprocating movement stroke of the pipe is 40 millimeters and the swing is 45°. The upper end of the pipe is open, and the test material is given from here. The lower part of the pipe is slender. A rubber tube with a clamp can be placed over its slender part. The clamp is used to adjust the water flow out. The flushing water is injected from the upper branch pipe of the glass tube.

3.Electric principle:
微信圖片_20250428151642.png 
4.Installation, and operation and matters needing attention to install:
XCGS - type magnetic separation tube host and excitation power just placed on the wooden frame, before use should be stripped all the wrapping paper, wipe rust-proof oil and check whether all the fastening screw loosening, It can be used only after confirming that the mechanical operation part and all the welding points of the part are in normal condition. On of the excitation and excitation power supplies and host the power of the motor, can boot commissioning (in the input voltage 220 v + 10 v, excitation current can be adjusted to the largest 4 a). Connect the glass tube with a rubber hose to fill water, and attach a soft rubber hose to the upper end of the glass tube.


Product Detail

Product Tags

Product Pictures
8004.jpg8002.jpg1.png

Product Parameters

NO

                           Item

Unit

   Numerical

               ; Note

1

 

Magnetic field

Interval

mm

52

 

 

 

Intensity(GS)continuously variable

MT

350

Stepless adjustable

2

Tube 

Diameter

mm

50

 

 

 

Vibrating freq

r/min

70

 

 

 

Shift motion

mm

40

 

3

Feed size

mm

0.5

 

4

Voltage

V

220

50Hz

5

Weight

kg

200